System Verification and Performance Tests
Performance Tests
14.Connect the short to Port 2.
15.Press REVERSE: SHORT, OPEN (F), DONE OPENS.
16.Connect the 50Ω termination to Port 2.
17. Press REVERSE: LOAD, SHORT (F), DONE SHORTS.
18.When the analyzer displays PRESS ‘DONE’ IF FINISHED WITH
STD(s), press STANDARDS DONE.
Wait for the message COMPUTING CAL COEFFICIENTS to disappear
from the analyzer’s display before proceeding to the next step.
19.Connect the equipment as shown in Figure 1-49.
Figure 1-49 Transmission Calibration Test Setup (Port 1 to Port 2)
20.Press TRANSMISSION, DO BOTH FWD + REV.
21.Press DONE 2-PORT CAL.