User`s guide
1-102 Chapter1
System Verification and Performance Tests
Performance Tests
Reflection Tracking (Reverse) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 40, x1, EXECUTE TEST.
2. When the test is done, press Scale Ref, AUTO SCALE, Marker,
MARKER 1
.
3. Using the front panel knob, locate the minimum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Press MARKER 2.
5. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
6. Press ∆ MENU MODE, ∆ REF=1. Write the value on the performance
test record.
7. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
8. Press Marker, All OFF.
Transmission Tracking (Forward) Calibration Coefficient
1. Press System, SERVICE MENU, TESTS, 37, x1, EXECUTE TEST.
2. When the analyzer finishes the test, press Scale Ref, AUTO SCALE,
Marker
.
3. Using the front panel knob, locate the minimum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
4. Press MARKER 2.
5. Using the front panel knob, locate the maximum value of the data
trace for the 300 kHz to 1.3 GHz frequency range.
6. Press ∆ MENU MODE, ∆ REF=1. Write the value on the performance
test record.
7. Repeat the previous two steps for the other frequency range(s) listed
on the performance test record.
8. Press Marker, All OFF.