User`s guide
Chapter 1 1-105
System Verification and Performance Tests
Performance Tests
Required Equipment
Analyzer warmup time: 30 minutes
Description Model Number or Part Number
Power Meter 436A/437B/438A or
E4418B/E4419B
Power Sensor 8482A
Step Attenuator, 110 dB
(See notes on the following page.)
8496A Option 003, H18
Adapter (2), APC-7 to Type-N (f) 11525A
Cable, GPIB 10833A
Calibration Kit, 7-mm, 50Ω 85031B
Attenuator, 20 dB 8492A Option 020
Cable, APC-7, 50Ω part number 8120-4781