User`s guide

Chapter 1 1-107
System Verification and Performance Tests
Performance Tests
A/R Trace Noise Phase IF BW = 10 Hz
1. Press Format, PHASE.
2. Press Sweep Setup, TRIGGER MENU, SINGLE. Wait for the sweep to
finish as indicated by the Hld notation on the left side of the display.
3. Press Scale Ref, AUTO SCALE.
4. Write the s.dev (standard deviation) value shown that appears on the
analyzer display on the performance test record.
A/R Trace Noise Phase IF BW = 3 kHz
1. Press Avg, IF BW, 3, k/m.
2. Press Sweep Setup, TRIGGER MENU, SINGLE. Wait for the sweep to
finish as indicated by the Hld notation on the left side of the display.
3. Press Scale Ref, AUTO SCALE.
4. Write the s.dev (standard deviation) value shown that appears on the
analyzer display on the performance test record.
A/R Trace Noise Magnitude IF BW = 3 kHz
1. Press Format, LOG MAG.
2. Press Sweep Setup, TRIGGER MENU, SINGLE. Wait for the sweep to
finish as indicated by the Hld notation on the left side of the display.
3. Press Scale Ref, AUTO SCALE.
4. Write the s.dev (standard deviation) value shown that appears on the
analyzer display on the performance test record.
B/R Trace Noise Magnitude IF BW = 3 kHz
1. Press Meas, Trans: FWD S31 (B/R), Sweep Setup, TRIGGER MENU,
SINGLE
. Wait for the sweep to finish as indicated by the Hld notation
on the left side of the display
2. Press Scale Ref, AUTO SCALE.
3. Write the s.dev (standard deviation) value shown that appears on the
analyzer display on the performance test record.