User`s guide
1-108 Chapter1
System Verification and Performance Tests
Performance Tests
B/R Trace Noise Phase IF BW = 3 kHz
1. Press Format, PHASE, Sweep Setup, TRIGGER MENU, SINGLE. Wait for
the sweep to finish as indicated by the Hld notation on the left side of
the display.
2. Press Scale Ref, AUTO SCALE.
3. Write the s.dev (standard deviation) value shown that appears on the
analyzer display on the performance test record.
B/R Trace Noise Phase IF BW = 10 Hz
1. Press Avg, IF BW, 10, x1, Sweep Setup, TRIGGER MENU, SINGLE. Wait for
the sweep to finish as indicated by the Hld notation on the left side of
the display.
2. Press Scale Ref, AUTO SCALE.
3. Write the s.dev value that appears on the analyzer display on the
performance test record.
B/R Trace Noise Magnitude IF BW = 10 Hz
1. Press Format, LOG MAG, Sweep Setup, TRIGGER MENU, SINGLE. Wait
for the sweep to finish as indicated by the Hld notation on the left
side of the display.
2. Press Scale Ref, AUTO SCALE.
3. Write the s.dev (standard deviation) value shown that appears on the
analyzer display on the performance test record.
Option 006:
1. Press Sweep Setup, CW FREQ, 6.
2. Go back to “A/R Trace Noise Magnitude IF BW = 10 Hz” on
page 1-106 and performthis entire procedure again. Record all values
in the performance test record under “Frequency 6 GHz.”