User`s guide
1-110 Chapter1
System Verification and Performance Tests
Performance Tests
11. Test Port Receiver Magnitude Dynamic Accuracy
Required Equipment
Analyzer warmup time: 30 minutes
NOTE The 8496A step attenuator (Option 003, H18) comes with a special
calibration that supports the measurement uncertainties expressed in
the test record for this performance test.
The special calibration consists of two measurements. The first is a
measurement of the attenuation at each step. The data reported for this
measurement have the following uncertainties:
• ±0.006 dB from 0 to 40 dB
• ±0.015 dB from >40 to 80 dB
• ±0.025 dB from >80 to 90 dB
• ±0.05 dB >90 dB
The second calibration measurement characterizes match stability
between attenuator settings for each attenuator port. The vector
difference of S11 or S22 between the reference attenuation step and all
the other steps is measured. The magnitude of this difference is certified
to be <0.0316 (>30 dB).
Description Model Number or Part Number
Power Meter 436A/437B/438A or
E4418B/E4419B
Power Sensor 8482A
Step Attenuator, 110 dB
See notes on the following page.
8496A Option 003, H18
Adapter (2), APC-7 to Type-N (f) 11525A
Cable, GPIB 10833A
Calibration Kit, 7-mm, 50Ω 85031B
Attenuator, 20 dB 8492A Option 020
Cable, APC-7, 50Ω part number 8120-4779