User`s guide
1-112 Chapter1
System Verification and Performance Tests
Performance Tests
3. In Table 1-3, transfer the 10 dB error value located within the
parenthesis in column B to each space in column C.
4. The values in column D result from changing the reference
attenuation of the calibration data of the 8496A to 10 dB.
Calculate the attenuation error values for this column by subtracting
the values in column C from the values in column B (B − C = D).
5. The values in column F result from correcting the expected
measurement value by the amount of attenuator error.
Calculate the attenuation error values for this column by subtracting
the values in column D from the values in column E (E − D = F).
6. Transfer the values from column F in Table 1-3 to column F in the
performance test record for both ports.
Table 1-3 Magnitude Dynamic Accuracy Calculations
A B C D: (B−C) E F: (E−D)
8496A Attn.
(dB)
Attn. Error
(ref 0 dB)
10 dB Error
Value
Attn. Error
(ref 10 dB)
Expected
Measurement
(dB)
Expected
Measurement
(corrected) (dB)
0 0 dB 10
10 (...............) 0 dB 0
20 −10
30 −20
40 −30
50 −40
60 −50
70 −60
80 −70
90 −80