User`s guide
1-114 Chapter1
System Verification and Performance Tests
Performance Tests
4. Set up the analyzer for power meter calibration:
a. Select the analyzer as the system controller:
• Local
• SYSTEM CONTROLLER
b. Set the power meter’s address:
• SET ADDRESSES
• ADDRESS:P MTR/GPIB, 13, x1
c. Select the appropriate power meter by pressing POWER MTR until
the correct model number is displayed (436A, 437B, 438A).
d. Select the power meter cal and enter the power sensor calibration
data by pressing:
• Cal, CAL KIT, SELECT CAL KIT, 7mm 85031
• Cal, PWRMTR CAL, LOSS/SENSOR LISTS,
CAL FACTOR SENSOR A
(enter the power sensor calibration
data for 30 MHz), DONE
5. Take a power meter calibration sweep.
• Cal, PWRMTR CAL, –20, x1
• ONE SWEEP, TAKE CAL SWEEP
6. Verify that the power meter reads approximately –20 dBm.