User`s guide

Chapter 1 1-115
System Verification and Performance Tests
Performance Tests
Full 2-Port Calibration
1. Connect the equipment as shown in Figure 1-52.
Figure 1-52 Full 2-Port Calibration
2. Perform a full 2-port error correction with isolation using the
85031B calibration kit.
NOTE When you are performing error-correction for a system that has type-N
port connectors, the softkey menus label the sex of the port
connector—not the calibration standard connector. For example, the
label SHORT (F) refers to the short connected to the female port.