User`s guide

1-118 Chapter1
System Verification and Performance Tests
Performance Tests
R-Channel Magnitude Dynamic Accuracy
1. Press Preset.
2. Connect the equipment as shown in Figure 1-54.
Figure 1-54 R–Channel Magnitude Dynamic Accuracy Test Setup
3. Press Avg, IF BW, 10, x1.
4. Press Sweep Setup, NUMBER of POINTS, 51, x1, CW FREQ, 30, M/µ,
SWEEP TYPE MENU, POWER SWEEP
.
5. Press Meas, INPUT PORTS, B.
6. Press Marker, –15, x1.
7. Press Marker, MARKER 2, –10, x1.
8. Press MARKER 3, –5, x1.
9. Press MARKER 4, 0, x1.
10.Press MARKER 5, 5, x1.
11.Press MODE MENU, REF = 2.
12.Press Chan 2, Meas, INPUT PORTS, R.
13.Press Display, DUAL | QUAD SETUP, DUAL CHAN ON.
14.Press Sweep Setup, TRIGGER MENU, SINGLE.
15.Press Display, MORE, D2/D1 TO D2 ON.