User`s guide

1-120 Chapter1
System Verification and Performance Tests
Performance Tests
12. Test Port Receiver Magnitude Compression
Perform this test to verify the magnitude compression/expansion of the
analyzer’s test port receivers.
Specifications
Required Equipment
Analyzer warmup time: 30 minutes
Frequency Range
Magnitude
a
a. With a 10 Hz IF bandwidth
300 kHz to 3 GHz 0.45 dB
3 GHz to 6 GHz
b
b. Option 006 only
0.80 dB
Description Model Number or Part Number
Cable (2), 50, APC-7 part number 8120-4779
Step Attenuator 8496A Option 003, Option H18