User`s guide

Chapter 1 1-121
System Verification and Performance Tests
Performance Tests
Procedure
1. Connect the equipment as shown in Figure 1-55.
Figure 1-55 Port Receiver Magnitude Compression Test Setup
2. On the analyzer, press Preset, Avg, IF BW, 10, x1.
3. Press Meas, INPUT PORTS, B.
4. Press Sweep Setup, SWEEP TYPE MENU, POWER SWEEP, Start, –10, x1.
5. Press Sweep Setup, CW FREQ, 50, M/µ.
6. Set the step attenuator to 20 dB attenuation.
7. Press TRIGGER MENU, SINGLE.
8. Wait for the sweep to finish, then press Display, DATA MEM,
DATA/MEM
.
9. Set the step attenuator to 0 dB.
10.Press Sweep Setup, TRIGGER MENU, SINGLE.
11.Press Scale Ref, AUTO SCALE.