User`s guide

1-132 Chapter1
System Verification and Performance Tests
Performance Tests
15. Harmonic Measurement Accuracy
Perform this test to verify the analyzer’s accuracy when operating in the
harmonic measurement mode (Option 002). The analyzer’s reading is
compared to that of a power meter. The allowable difference (accuracy) is
shown in the “Specifications” table.
Specifications
Required Equipment
Analyzer warmup time: 30 minutes
Frequency Range Accuracy
16 MHz to 3 GHz 1.5 dB
3 GHz to 6 GHz
a
3.0 dB
a. Only for analyzers with Option 006
Description Model Number or Part Number
External Source 83712B
Power Meter 436A/437B/438A or
4418B/4419B
Power Splitter 11667A Option 001
Power Sensor 8482A
Adapter, Type-N (f) to Type-N (f) part number 1250-0777
Cable (2) 50, Type-N, 24-inch part number 8120-4781
Cable, BNC, 24-inch part number 8120-1840
Additional equipment needed for analyzers with Option 006:
Power Sensor 8481A