User`s guide
2-10 Chapter2
Operation
Operation
Example: Duplexer Setup
The following example shows how to set up the analyzer to measure from
Port 1 to Port 3 on channels 1 and 3, and how to measure from Port 3 to
Port 2 on channels 2 and 4. In this example, steps 4 and 7 indicate where
to perform a full 2-port calibration. Refer to the standard documentation
for details outlining how to perform a full 2-port calibration. This
example assumes the analyzer has been configured so that the default
state at instrument power up or preset is in 3 PORT mode.
1. Press Preset.
2. Turn the coupled channel off by pressing Sweep Setup and toggling
COUPLED CH from on to OFF.
3. Press Chan 1.
4. Perform a full 2-port calibration for Port 1 and Port 3. The
S-parameters displayed during the calibration indicate the ports
selected. After the calibration is complete, save the calibration by
pressing Save/Recall, Save State.
5. Press Chan 2, Meas. Toggle SELECT PORTS [1-3] to SELECT PORTS [2-3]
by pressing the SELECT PORTS key.
6. Perform a full 2-port calibration for Port 2 and Port 3. The
S-parameters displayed during the calibration indicate the ports
selected. After the calibration is complete, save the calibration by
pressing Save/Recall, Save State.
7. Turn on the quad channel by pressing Display, DUAL | QUAD SETUP,
4 PARM DISPLAY. Select SETUP A.