User`s guide

Chapter 1 1-133
System Verification and Performance Tests
Performance Tests
Determine the Cable Loss at Specified Frequencies
1. Connect the equipment as shown in Figure 1-59.
Figure 1-59 Harmonic Measurement Accuracy, Cable Loss Setup 1
2. Press Preset, Start, 40, M/µ.
3. Press Meas, Trans: S31 (B/R), Cal, CALIBRATE MENU, RESPONSE, THRU.