User`s guide
2-4 Chapter2
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
1. Test Port Output Frequency Range and Accuracy
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
CW Frequency
(MHz)
Min.
(MHz)
Results Measured
(MHz)
Max.
(MHz)
Meas. Uncert.
(MHz)
0.03 0.029 999 7 0.030 000 3 ±0.000 000 050
0.3 0.299 997 0.300 003 ±0.000 000 520
5.0 4.999 950 5.000 050 ±0.000 008 610
16.0 15.999 840 16.000 160 ±0.000 028 220
31.0 30.999 690 31.000 310 ±0.000 0530 730
60.999 999 60.999 390 61.000 610 ±0.000 104 800
121.0 120.998 790 121.001 210 ±0.000 206 800
180.0 179.998 200 180.001 800 ±0.000 307 200
310.0 309.995 900 310.003 100 0.000 528 300
700.0 699.930 000 700.007 000 ±0.001 191 700
1 300.0 1 299.987 1 300.013 ±0.002 212 300
2 000.0 1 999.980 2 000.020 ±0.003 403 000
3 000.0 2 999.970 3 000.030 ±0.005 104 000
4 000.0 3 999.960 4 000.040 ±0.006 805 000
5 000.0 4 999.950 5 000.050 ±0.008 506 000
6 000.0 5 999.940 6 000.060 ±0.010 207 000