User`s guide
2-6 Chapter2
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
3. Test Port Output Power Level Accuracy
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Test Frequency Test Port
Output Power
(dBm)
Specification
(dB)
Measured Value
(dBm)
Measurement
Uncertainty
(dB)
300 kHz 0 ±1 ±0.47
20 MHz 0 ±1 ±0.14
50 MHz 0 ±1 ±0.14
100 MHz 0 ±1 ±0.14
200 MHz 0 ±1 ±0.14
500 MHz 0 ±1 ±0.14
1 GHz 0 ±1 ±0.13
2 GHz 0 ±1 ±0.13
3 GHz 0 ±1 ±0.27
4 GHz 0 ±1 ±0.17
5 GHz 0 ±1 ±0.17
6 GHz 0 ±1 ±0.17