User`s guide

2-8 Chapter2
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
4. Test Port Output Power Linearity (continued)
Frequency Range: 3 GHz to 6 GHz
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Power Setting
dBm
Marker 1
Maximum
(dB)
Marker 2
Minimum (dB)
Specification
(dB)
Meas.
Uncert.
(dB)
15 ±0.2 ±0.04
13 ±0.2 ±0.03
11 ±0.2 ±0.03
9 ±0.2 ±0.03
7 ±0.2 ±0.03
–5 ±0.2 ±0.03
–3 ±0.2 ±0.02
–1 ±0.2 ±0.02
+1 ±0.2 ±0.02
+3 ±0.2 ±0.02
+5 ±0.2 ±0.03
+7 ±0.5 ±0.03
+8 ±0.5 ±0.03
+9 ±0.5 ±0.03
+10 ±0.5 ±0.03