User`s guide

2-12 Chapter2
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
8. Test Port Crosstalk
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Test Settings Specification
(dB)
Measured Value
(dB)
Meas.
Uncert. (dB)
Port 1 and Port 3
Crosstalk to Test Port 3
±1.0
300 kHz – 3 GHz < –100
Crosstalk to Test Port 1
±1.0
300 kHz – 3 GHz < –100
Crosstalk to Test Port 1
±2.0
3 GHz – 6 GHz < –87
Crosstalk to Test Port 3
±2.0
3 GHz – 6 GHz < –87
Port 2 and Port 3
Crosstalk to Test Port 3
±1.0
300 kHz – 3 GHz < –100
Crosstalk to Test Port 2
±1.0
300 kHz – 3 GHz < –100
Crosstalk to Test Port 2
±2.0
3 GHz – 6 GHz < –87
Crosstalk to Test Port 3
±2.0
3 GHz – 6 GHz < –87
Port 1 and Port 2
Crosstalk to Test Port 2
±1.0
300 kHz – 3 GHz < –100
Crosstalk to Test Port 1
±1.0
300 kHz – GHz < –100
Crosstalk to Test Port 1
±2.0
3 GHz – 6 GHz < –87
Crosstalk to Test Port 2
±2.0
3 GHz – 6 GHz < –87