User`s guide
Chapter 2 2-13
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
9. Uncorrected Port Performance, Port 1 and Port 3
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Test Description Frequency Range Spec.
(dB)
Measured Value
(dB)
Meas.Uncert.
(dB)
Forward Direction
Directivity 300 kHz – 1.3 GHz ≤35 ±0.9
1.3 GHz – 3 GHz ≤30 ±0.8
3 GHz – 6 GHz ≤20 ±0.8
Forward Direction
Source Match 300 kHz – 1.3 GHz ≤16 ±0.2
1.3 GHz – 3 GHz ≤16 ±0.2
3 GHz – 6 GHz ≤12 ±0.3
Reverse Direction
Trans. Tracking 300 kHz – 1.3 GHz ±1.5 ±0.006
1.3 GHz – 3 GHz ±1.5 ±0.009
3 GHz – 6 GHz ±2.5 ±0.021
Forward Direction
Refl. Tracking 300 kHz – 1.3 GHz ±1.5 ±0.001
1.3 GHz – 3 GHz ±1.5 ±0.005
3 GHz – 6 GHz ±2.5 ±0.020
Reverse Direction
Load Match 300 kHz – 1.3 GHz ≤18 ±0.1
1.3 GHz – 3 GHz ≤16 ±0.2
3 GHz – 6 GHz ≤14 ±0.2