User`s guide

2-14 Chapter2
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
9. Uncorrected Port Performance, Port 1 and Port 3 (continued)
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Test Description Frequency Range Spec.
(dB)
Measured Value
(dB)
Meas. Uncert.
(dB)
Forward Direction
Load Match 300 kHz 1.3 GHz 18 ±0.1
1.3 GHz 3 GHz 16 ±0.2
3 GHz 6 GHz 14 ±0.2
Reverse Direction
Directivity 300 kHz 1.3 GHz 35 ±0.9
1.3 GHz 3 GHz 30 ±0.8
3 GHz 6 GHz 20 ±0.8
Reverse Direction
Source Match 300 kHz 1.3 GHz 16 ±0.02
1.3 GHz 3 GHz 16 ±0.02
3 GHz 6 GHz 12 ±0.03
Reverse Direction
Refl. Tracking 300 kHz 1.3 GHz ±1.5 ±0.001
1.3 GHz 3 GHz ±1.5 ±0.005
3 GHz 6 GHz ±2.5 ±0.020
Forward Direction
Trans. Tracking 300 kHz 1.3 GHz ±1.5 ±0.006
1.3 GHz 3 GHz ±1.5 ±0.009
3 GHz 6 GHz ±2.5 ±0.021