User`s guide

Chapter 2 2-21
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
11. Test Port Receiver Magnitude Dynamic Accuracy (Test Port 1)
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
GF|GF|
Test Port
Input
Power
(dBm)
8496A
Attn.
(dB)
Test Port
Measurement
(dB)
Expected
Measurement
(Corrected)
(dB)
Dynamic
Accuracy
(Calculated)
Spec.
(dB)
Meas.
Uncert.
(dB)
Test Port 1
–10 0 0.033 0.008
–20 (ref) 10 0.000 0.020 0.008
–30 20 0.031 0.008
–40 30 0.041 0.008
–50 40 0.054 0.008
–60 50 0.069 0.017
–70 60 0.111 0.017
–80 70 0.257 0.017
–90 80 0.741 0.017
–100 90 2.121 0.027