User`s guide
Chapter 2 2-21
Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
11. Test Port Receiver Magnitude Dynamic Accuracy (Test Port 1)
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
GF|G – F|
Test Port
Input
Power
(dBm)
8496A
Attn.
(dB)
Test Port
Measurement
(dB)
Expected
Measurement
(Corrected)
(dB)
Dynamic
Accuracy
(Calculated)
Spec.
(dB)
Meas.
Uncert.
(dB)
Test Port 1
–10 0 ≤0.033 0.008
–20 (ref) 10 0.000 ≤0.020 0.008
–30 20 ≤0.031 0.008
–40 30 ≤0.041 0.008
–50 40 ≤0.054 0.008
–60 50 ≤0.069 0.017
–70 60 ≤0.111 0.017
–80 70 ≤0.257 0.017
–90 80 ≤0.741 0.017
–100 90 ≤2.121 0.027