Performance Test Records
Option H39 Performance Test Record
Model 8753ES Option H39
Serial Number
Report Number
Option(s) Date
15. Harmonic Measurement Accuracy (Option 002 only)
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Analyzer
Freq.
Ext.
Source
Freq.
Power
Meter
Meas.
Cable
Loss
(A)
Power
Meter +
Cable
Loss
(B)
Network
Analyzer
Reading
A–B Spec.
(dB)
Meas.
Uncert.
(dB)
20 MHz 40 MHz 1.5 ±0.17
20 MHz 60 MHz 1.5 ±0.17
100 MHz 200 MHz 1.5 ±0.17
100 MHz 300 MHz 1.5 ±0.17
300 MHz 600 MHz 1.5 ±0.17
300 MHz 900 MHz 1.5 ±0.17
500 MHz 1 GHz 1.5 ±0.17
500 MHz 1 GHz 1.5 ±0.17
1 GHz 2 GHz 1.5 ±0.17
1 GHz 2 GHz 1.5 ±0.17
2 GHz 4 GHz 3.0 ±0.17
2 GHz 6 GHz 3.0 ±0.17
3 GHz 6 GHz 3.0 ±0.17