User`s guide
3-6 Chapter3
Theory of Operation and Troubleshooting
Troubleshooting
Troubleshooting
This section only covers troubleshooting techniques for the three-port
test set with the 3PORT mode off. To turn the 3PORT mode off, press
System, CONFIGURE MENU, OPTION HXX CONFIG, 3PORT MODE and toggle
to OFF. For information on troubleshooting other portions of the analyzer,
refer to the standard documentation.
Required Equipment
Description Model Number or Part Number
Calibration Kit, 50Ω Type-N 85032B
Power Meter 436A, 437A, 438A
Sensor 8481A
RF Cable 50Ω Type-N (m) to (m) part number 8120-4781
Table 3-1 Test Set I/O Port Parameters
Measurements Ports and Port
Paths
S-parameters
Under Meas
Menu
Test Set
I/O Setting
Test Set
I/O Setting
FWD REV
Reflection 1) Port 1 S
11
7 ---
2) Port 2 S
22
65
3) Port 3 S
33
--- 7
Transmission 4) Port 1-Port 3 S
31
, S
13
77
5) Port 2-Port 3 S
32
, S
23
66
6) Port 1-Port 2 S
21
, S
12
55