Test Set I/O Port Parameters
Table 2-4 shows the nine S-parameters that can be measured with the
analyzer.
Port 1, Port 2, and Port 3 can be switched independently depending on
the test set I/O forward (FWD) or reverse (REV) value and the analyzer’s
test state FWD or REV. When the analyzer is in the dual-channel mode
and the channels are uncoupled, the test set I/O (FWD and REV) can be
set differently regardless of the FWD and REV I/O setting. The
S-parameters are updated on the display when either Chan 1 or Chan 2 is
pressed on the active channel front panel keys. S-parameters can also be
updated through the Meas menu. After pressing the Meas key, select an
S-parameter value (for example Refl: FWD S11 A/R) to update the display.
CAUTION Before performing a full 2-port calibration, make sure that both the
TESTSET I/O FWD and the TESTSET I/O REV settings are set the
same.