User`s guide

4-2 Chapter4
Specifications
Option H39 Specifications
Option H39 Specifications
All specifications for the Option H39 are consistent with those for the
standard analyzer with the exception of the specifications found in this
chapter.
Measurement Capabilities
Frequency Response
Port Paths S-Parameter Measurements
Port 1 to Port 3
a
a. Default for preset and power up
S
11
, S
21
, S
12
, S
22
Port 1 to Port 2 S
11
, S
21
, S
12
, S
22
Port 2 to Port 3 S
11
, S
21
, S
12
, S
22
Frequency Response
a
a. At 25 ±5 ˚C
Specification
Port 1 and Port 3
300 kHz to 3 GHz ±1 dB
3 GHz to 6 GHz ±2 dB
Port 2
b
b. Peak-to-peak (p-p) deviation from nominal trace
across the frequency range.
300 kHz to 3 GHz 2 dB (p-p)
3 GHz to 6 GHz 4 dB (p-p)