User`s guide

1-6 Chapter1
System Verification and Performance Tests
Introduction
General Instrument Setup Information
In the verification procedure, the analyzer is set up using type-N (m) to
APC-7 adapters. The adapters are installed on Port 1 and Port 3. Cables
are then connected to the adapters. This setup is used throughout the
verification procedure. Calibration of the instrument is done with the
adapters and cables attached. If the cables are removed, the calibration
will be affected and the performance of the instrument will be invalid.
All connections to Port 1 and Port 3 are done at the open end of the cable.
The analyzer will be set up to use the turn-on and preset default ports
(Port 1 and Port 3). System verification of the analyzer requires less
setup by leaving the adapters and cables on the analyzer and utilizing
the default settings.
NOTE Once calibration is completed with the adapters and cables attached to
the analyzer, the adapters and cables should remain on the analyzer
until all of the verification procedure has been completed. If the adapters
or cables are removed the calibration is affected and the performance of
the instrument will be invalid.