User`s guide

Chapter 1 1-9
System Verification and Performance Tests
System Verification
Port 1 to Port 3 Initialization
Instrument Setup
CAUTION The following procedure will erase all instrument states that may be
stored in internal memory.
1. Perform the following steps to save to a floppy disk any instrument
states that are stored in internal memory.
a. Press Save/Recall, SELECT DISK, INTERNAL MEMORY, RETURN.
b. Select an instrument state and press RECALL STATE.
c. Press SELECT DISK, INTERNAL DISK, RETURN, SAVE STATE.
d. If the instrument state file was not saved to disk with the same
name that it had while in internal memory, you may wish to
rename the file.
2. Clear all internal memory. Press System, SERVICE MENU, PEEK/POKE,
RESET MEMORY, Preset
.
3. Connect the following equipment to Port 1 and Port 3 of the analyzer
as shown in Figure 1-1.
Analyzer warmup time: 30 minutes
Adapter (2), APC-7 to Type-N (m) 11525A
Port Extension Cable Set, 7-mm 11857D