User`s guide

1-10 Chapter1
System Verification and Performance Tests
System Verification
Figure 1-1 System Verification Test Setup
1. While the equipment is warming up, review “Principles of Microwave
Connector Care” in the standard 8753ES service guide. Good
connections and clean, undamaged connectors are critical for accurate
measurement results.
2. Insert the verification kit disk into the analyzer disk drive.
3. Press Preset, Save/Recall, SELECT DISK, INTERNAL DISK.
4. If you want a printout of the verification data for all the devices, press
System, SERVICE MENU, TEST OPTIONS, RECORD ON.
CAUTION If you switch on the record function, you cannot switch it off during the
verification procedure.
5. Position the paper in the printer so printing starts at the top of the
page.