User`s guide
1-12 Chapter1
System Verification and Performance Tests
System Verification
Port 1 to Port 3 Measurement Calibration
1. Press Cal, CAL KIT, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN,
CALIBRATE MENU, FULL 2-PORT.
2. Press ISOLATION, OMIT ISOLATION.
3. Press REFLECTION.
4. Connect the open to reference test port 1 as shown in Figure 1-2.
Figure 1-2 Connections for Measurement Calibration Standards
5. Press FORWARD:OPEN.
a. When the analyzer finishes measuring the standard, connect the
short to reference test port 1.
b. Press FORWARD:SHORT.
c. When the analyzer finishes measuring the standard, connect the
50Ω termination (supplied in the calibration kit) to reference test
port 1.
d. Press FORWARD:LOAD