User`s guide
Chapter 1 1-13
System Verification and Performance Tests
System Verification
6. When the analyzer finishes measuring the standard, connect the open
to reference test port 3.
a. Press REVERSE:OPEN.
7. When the analyzer finishes measuring the standard, connect the
short to reference test port 3.
a. Press REVERSE:SHORT.
b. When the analyzer finishes measuring the standard, connect the
50Ω termination to reference test port 3.
c. Press REVERSE:LOAD.
d. When the analyzer finishes measuring the standard, press
STANDARDS DONE.
8. The analyzer should briefly display COMPUTING CAL
COEFFICIENTS.
Figure 1-3 Transmission Calibration Setup
9. Connect the port cables as shown in Figure 1-3.
10.Press TRANSMISSION, DO BOTH FWD + REV.
11.Press DONE 2-PORT CAL.