System Verification and Performance Tests
System Verification
12.Press Save/Recall, SELECT DISK, INTERNAL MEMORY, RETURN,
SAVE STATE
to save the calibration in the analyzer internal memory.
13.When the analyzer finishes saving the instrument state, press
SELECT DISK, INTERNAL DISK.
Port 1 to Port 3 Device Verification
1. Press System, SERVICE MENU, TESTS 28, x1, EXECUTE TEST.
2. At the prompt, connect the 20 dB attenuator (supplied in the
verification kit) as shown in Figure 1-4.
3. Press CONTINUE to run the test:
a. If you switched off the record function, you must press CONTINUE
after each S-parameter measurement.
b. If you switched on the record function, the analyzer measures all
S-parameters (magnitude and phase) without pausing. The
analyzer only displays and prints the PASS/FAIL information for
the S-parameter measurements that are valid for system
verification.
Figure 1-4 Connections for the 20 dB Verification Test