User`s guide
1-16 Chapter1
System Verification and Performance Tests
System Verification
6. When all measurements are complete, replace the verification device
with the verification mismatch, as shown in Figure 1-6. Be sure to
connect Port A of the verification mismatch to Port 3.
Figure 1-6 Mismatch Device Verification Setup 1
7. Press RETURN, TESTS, 30, x1, EXECUTE TEST, CONTINUE.