User`s guide
Chapter 1 1-21
System Verification and Performance Tests
Performance Tests
1. Test Port Output Frequency Range and Accuracy
Perform this test to verify the frequency accuracy of the analyzer over its
entire operating frequency range.
Specifications
Required Equipment
Analyzer warmup time: 30 minutes
Frequency Range
Frequency Accuracy
a
30 kHz to 3 GHz ± 10 ppm
3 GHz to 6 GHz
b
± 10 ppm
a. At 25 ± 5 °C
b. Only for analyzers with Option 006
Description Model Number or Part Number
Frequency Counter 5350B/51B52B
Cable, 50Ω Type-N, 24-inch part number 8120-4781
Adapter, APC-2.5 (f) to Type-N (f) part number 1250-1745
Adapter, Type-N (f) to BNC (m) part number 1250-1477