3. Test Port Output Power Level Accuracy
Perform this test to confirm the accuracy of the 8753ES test port output
power. A power meter is used to determine the output level from Port 1.
This measurement is compared to the level set by the network analyzer.
The difference must be within the specified tolerance.
Analyzer warmup time: 30 minutes
Specifications
Required Equipment
Frequency Range Test Port Output
Power Accuracy
a
a. At 0 dBm and 25° C ±5° C
300 kHz to 3 GHz ±1.0 dB
3 GHz to 6 GHz
b
b. Only for analyzers with Option 006
±1.0 dB
Description
Model Number
Power Meter 436A/437B/438A or E4418B/4419B
Power Sensor 8482A
Additional Equipment for Option 006 Analyzers
Power Sensor 8481A