User`s guide
1-30 Chapter1
System Verification and Performance Tests
Performance Tests
4. Test Port Output Power Linearity
Specifications
Required Equipment
This test verifies the analyzer’s port output power linearity throughout
the analyzer’s frequency range. A power meter calibration is used to set
the reference output power. Subsequent measurements are made with
the analyzer’s Port 2 receiver.
Power Range
Power Level Linearity
a
a. Relative to 0 dBm from 300 kHz to 3 GHz
(6 GHz for Option 006).
−15 to +5 dBm ±0.2 dB
+5 to +10 dBm ±0.5 dB
+5 to +8 dBm ±0.5 dB
Description Model Number or Part Number
Power Meter 437B/438A or
E4418B/4419B
Power Sensor 8482A
Attenuator, APC-7, 50Ω, 20 dB 8492A Opt. 020
RF Cable, APC-7, 50Ω part number 8120-4781
GPIB Cable 10833A/B/C/D
Additional Equipment Required for Analyzers with Option 006
Power Sensor 8481A