1-32 Chapter1
System Verification and Performance Tests
Performance Tests
3. Press Preset.
4. Press Avg, IF BW, 100, x1, Sweep Setup, NUMBER of POINTS, 51, x1.
5. Press Start, 300, k/m then:
• If the analyzer has Option 006, press Stop, 3, G/n.
6. Press Local, SYSTEM CONTROLLER.
7. Press SET ADDRESSES and POWER MTR until the analyzer shows the
correct power meter model.
8. Press ADDRESS:P MTR/GPIB. The default power meter GPIB address
is 13. Make sure it is the same as your power meter GPIB address.
Otherwise, use the analyzer front panel keypad to enter the correct
GPIB address for your power meter.
9. Press Power, PWR RANGE AUTO, then toggle from man to MAN to turn
the auto power range off.
NOTE The analyzer displays the PRm annotation, indicating that the analyzer
power range is set to MANUAL.
10.Press PORT POWER to uncouple the port output power.
11.Press Cal, PWRMTR CAL.
12.Press LOSS/SENSR LISTS, CAL FACTOR SENSOR A. Refer to the back of
the power sensor to locate the different calibration factor values along
with their corresponding frequencies.
NOTE The analyzer’s calibration factor sensor table can hold a maximum of 55
calibration factor data points.