User`s guide
Chapter 1 1-33
System Verification and Performance Tests
Performance Tests
The following softkeys are included in the sensor calibration factor
entries menu:
The following keystrokes are used as an example for entering the first
two calibration factor data points for the 8482A power sensor (assuming
CF% = 96.4 at 100 kHz and CF% = 98.4 at 300 kHz):
a. From the sensor calibration factor entries menu, press ADD.
b. Press FREQUENCY, 100, k/m. If you make an entry error,
press ← and re-enter the correct value again.
c. Press CAL FACTOR, 96.4, x1.
d. Press DONE to terminate the first calibration factor data point
entry.
e. To enter the second cal factor data point, press ADD.
f. Press FREQUENCY, 300, k/m.
g. Press CAL FACTOR, 98.4, x1.
h. To terminate the second calibration factor data point entry, press
DONE.
i. Press SEGMENT and use the front panel knob to scroll through the
sensor calibration factors table. Check to be sure all values are
entered correctly. If you see an error, use the front panel knob to
point to the data point you want to modify and press EDIT.
SEGMENT Press to select a point where you can use the front panel
knob or entry keys to enter a value.
EDIT Press to edit or change a previously entered value.
DELETE Press to delete a point from the sensor calibration factor
table.
ADD Select this key to add a point into the sensor calibration
factor table.
CLEAR LIST Select this key to erase the entire sensor calibration factor
table.
DONE Select this key when done entering points to the sensor
calibration factor table.