1-34 Chapter1
System Verification and Performance Tests
Performance Tests
j. Press the appropriate softkeys and create a power sensor
calibration factors table.
k. Press DONE to exit the sensor calibration factor entries menu.
l. Press RETURN, ONE SWEEP, TAKE CAL SWEEP to start the power
meter calibration.
NOTE The analyzer displays the PC annotation indicating the power meter
calibration is done and the error correction is active.
Output Power Linearity Measurement: 300 kHz to 3 GHz
1. Set up the equipment as shown in Figure 1-12.
Figure 1-12 Port Output Power Level Linearity Test Setup
2. Press Meas, INPUT PORTS, B.
3. Press Sweep Setup, TRIGGER MENU, SINGLE.
4. Press Display, DATA → MEMORY, DATA/MEM.
5. Press Power, –15 (or the next power setting from the performance test
record), x1.
6. Press Sweep Setup, TRIGGER MENU, SINGLE.