User`s guide

1-38 Chapter1
System Verification and Performance Tests
Performance Tests
5. Minimum R Channel Level
Perform this test to determine the minimum R channel input power level
that phase lock can be accomplished.
Specifications
Required Equipment
Analyzer warmup time: 1 hour
Frequency Range Minimum R Channel Level
300 kHz to 3 GHz <–35 dBm
3 GHz to 6 GHz
a
a. Only for analyzers with Option 006
<–30 dBm
Description Model Number or Part Number
Cable, 50, Type-N, 24-inch part number 8120-4781
Adapter, APC 3.5 (m) to Type-N (f) part number 1250-1750