User`s guide
Chapter 1 1-39
System Verification and Performance Tests
Performance Tests
1. Connect the equipment as shown in Figure 1-14.
Figure 1-14 Minimum R Channel Level Test Setup
2. Press Preset, Meas, INPUT PORTS, R.
3. Press Power, PWR RANGE MAN, POWER RANGES, RANGE 5 –65 to –40.
4. Press Scale Ref, REFERENCE VALUE, –70, x1.
5. Press Sweep Setup, CW FREQ, 300, k/m.
6. Press Power, –65, x1.
The analyzer displays the message CAUTION: NO IF FOUND:
CHECK R INPUT LEVEL.
7. Press ↑ to increase the port power by 1 dBm.
8. If the analyzer displays a phase lock error message, continue
increasing the port power until phase lock is achieved.
9. Write the port power that is displayed on the analyzer on the
performance test record.
10.Repeat steps 5 through 9 for the other CW frequencies listed in the
performance test record.