User`s guide
Chapter 1 1-45
System Verification and Performance Tests
Performance Tests
Specifications
Required Equipment
Analyzer warmup time: 1 hour
Frequency Range Port IF Bandwidth Average Noise Level
300 kHz to 3.0 GHz Port 1 3 kHz −81 dBm
300 kHz to 3.0 GHz Port 1 10 Hz −102 dBm
300 kHz to 3.0 GHz
Port 2
a
3 kHz −81 dBm
300 kHz to 3.0 GHz
Port 2
a
10 Hz −102 dBm
300 kHz to 3.0 GHz Port 3 3 kHz −81 dBm
300 kHz to 3.0 GHz Port 3 10 Hz −102 dBm
3.0 GHz to 6.0 GHz
b
Port 1 3 kHz −76 dBm
3.0 GHz to 6.0 GHz
b
Port 1 10 Hz −97 dBm
3.0 GHz to 6.0 GHz
b
Port 2
a
3 kHz −76 dBm
3.0 GHz to 6.0 GHz
b
Port 2
a
10 Hz −97 dBm
3.0 GHz to 6.0 GHz
b
Port 3 3 kHz −76 dBm
3.0 GHz to 6.0 GHz
b
Port 3 10 Hz −97 dBm
a.A and B channel noise floor
b.Only for analyzers with Option 006
Description Model Number or Part Number
50Ω Type-N (m) Terminations (3) 909F, Option 012










