Specifications
5-6
Error Messages
Error Messages in Alphabetical Order
CANNOT MODIFY FACTORY PRESET
Error Number
199
You have attempted to rename, delete, or otherwise alter the factory preset
state. The factory preset state is permanently stored in CMOS memory and
cannot be altered. If your intent was to create a user preset state, you must
create a new instrument state, save it, and then rename it to "UPRESET". Refer
to Chapter 7 , “Preset State and Memory Allocation” for more detailed
instructions.
CANNOT READ/WRITE HFS FILE SYSTEM
Error Number
203
The disk is being accessed by the analyzer and is found to contain an HFS
(hierarchical file system) or files nested within subdirectories. The analyzer does
not support HFS. Replace the disk medium with a LIF or DOS formatted disk
that does not contain files nested within subdirectories.
CAN'T STORE/LOAD SEQUENCE, INSUFFICIENT MEMORY
Error Number
127
Your sequence transfer to or from a disk could not be completed due to
insufficient memory.
CAUTION: AUX CHANNELS MEASURE S-PARAMETERS ONLY
Error Number
216
This message appears if you try to assign a non-S-parameter measurement to
an auxiliary channel and a full 2-port cal is performed on channels 1 and 2 with
both channels uncoupled.
CAUTION: CORRECTION OFF: AUX CHANNEL(S) DISABLED
Error Number
215
This message is displayed when correction is forced off due to a stimulus change
that is not compatible with the current calibration while an auxiliary channel is
enabled. The auxiliary channels are restored when correction is turned on by
pressing .
CAUTION: FLOPPY DISK IS FULL
Error Number
218
This message is displayed if you try to save files to the floppy disk that does not
have enough memory to perform the save task. Correct by inserting anew floppy
disk in the front panel disk drive or by erasing files from the current floppy disk.
CAUTION: TOO MANY SEGMENTS OR POINTS
Information
Message
This message is displayed if you try to insert too many segments or points using
your current function.
CH1 (CH2, CH3, CH4) TARGET VALUE NOT FOUND
Error Number
159
Your target value for the marker search function does not exist on the current
data trace.
Cal
CORRECTION on OFF