Specifications

5-12
Error Messages
Error Messages in Alphabetical Order
ILLEGAL UNIT OR VOLUME NUMBER
Error Number
46
The disk unit or volume number set in the analyzer is not valid. Refer to the
disk drive operating manual.
INIT DISK removes all data from disk
Information
Message
Continuing with the initialize operation will destroy any data currently on the
disk.
INITIALIZATION FAILED
Error Number
47
The disk initialization failed, probably because the disk is damaged.
INSTRUMENT STATE MEMORY CLEARED
Error Number
56
All instrument state registers have been cleared from memory along with any
saved calibration data, memory traces, and calibration kit definitions.
Additionally, all user-settable selections (such as GPIB addresses) are set to
their defaults.
INSUFFICIENT MEMORY
Error Number
51
Your last front panel or GPIB request could not be implemented due to
insufficient memory space. In some cases, this is a fatal error from which you
can escape only by presetting the instrument.
INSUFFICIENT MEMORY FOR PRINT/PLOT
Error Number
168
There is not enough memoryavailable for the printor plot function. Increase the
available memory by changing or eliminating a memory-intensive operation
such as reducing the number of points in the sweep.
INSUFFICIENT MEMORY, PWR MTR CAL OFF
Error Number
154
There is not enough memory space for the power meter calibration array.
Increase the available memory by clearing one or more save/recall registers, or
by reducing the number of points.
INVALID KEY
Error Number
2
You pressed an undefined softkey.
LIMIT TABLE EMPTY
Error Number
205
Limit lines cannot be turned on unless a limit table has been created. Refer to
the “Making Measurements” chapter of the user’s guide for information on how
to create a limit table.