Error Messages
Error Messages in Alphabetical Order
LIST MODE OFF: INVALID WITH LO FREQ
Error Number
182
List mode has been turned off in the frequency offset mode because it is
incompatible with your selected LO frequency.
LIST TABLE EMPTY
Error Number
9
The frequency list is empty. To implement list frequency mode, add segments to
the list table.
LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN
Error Number
150
A logarithmic sweep is only valid if the stop frequency is greater than four times
the start frequency. For frequency spans of less than two octaves, the sweep type
automatically reverts to linear sweep.
LOW PASS: FREQ LIMITS CHANGED
Information
Message
The frequency domain data points must be harmonically related from dc to the
stop frequency. That is, stop = n × start, where n = number of points. If this
condition is not true when a low pass mode (step or impulse) is selected and
transform is turned on, the analyzer resets the start and stop frequencies. The
stop frequency is set close to the entered stop frequency, and the start frequency
is set equal to stop/n.
MEMORY FOR CURRENT SEQUENCE IS FULL
Error Number
132
All the memory in the sequence you are modifying is filled with instrument
commands.
MORE SLIDES NEEDED
Error Number
71
When you use a sliding load (in a user-defined calibration kit), you must set at
least three slide positions to complete the calibration.
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number
69
The softkey is not valid unless a calibration is
already in progress. Start a new calibration.
NO DISK MEDIUM IN DRIVE
Error Number
41
You have no disk in the current disk unit. Insert a disk, or check the disk unit
number stored in the analyzer.
NO FAIL FOUND
Service Error
Number 114
The self-diagnose function of the instrumentoperates on an internaltest failure.
At this time, no failure has been detected.