Specifications
5-22
Error Messages
Error Messages in Alphabetical Order
SYST CTRL OR PASS CTRL IN LOCAL MENU
Error Number
36
The analyzer is in talker/listener mode. In this mode, the analyzer cannot
control a peripheral device on the bus. Use the local menu to change to system
controller or pass control mode.
TEST ABORTED
Error Number
113
You have prematurely stopped a service test.
THIS LIST FREQ INVALID IN HARM/3 GHZ RNG
Error Number
133
You have set frequencies in the list that are outside of the allowable frequency
range for harmonic measurements, or are greater than 3 GHz on instruments
without Option 006. Reduce the frequency range of the list.
TOO MANY NESTED SEQUENCES. SEQ ABORTED
Error Number
164
You can only nest sequences to a maximum level of six. The sequence will abort
if you nest more than six.
TOO MANY SEGMENTS OR POINTS
Error Number
50
You can have a maximum of 30 segments or 1601 points in frequency list mode.
In power meter calibrations, you can have a maximum of 12 segments for power
sensor cal factors and power loss functions.
TRANSFORM, GATE NOT ALLOWED
Error Number
16
You can perform a time domain transformation only in linear and CW sweep
types.
TROUBLE! CHECK SETUP AND START OVER
Service Error
Number 115
Your equipment setup for the adjustment procedure in progress is not correct.
Check the setup diagram and instructions in the service guide. Start the
procedure again.
WAITING FOR CLEAN SWEEP
Information
Message
In single sweep mode, the instrument ensures that all changes to the
instrument state, if any, have been implemented before taking the sweep. The
command that you have initiated is being processed and will not be complete
until the new sweep is completed. An asterisk * is displayed in the left margin
until a complete fresh sweep has been taken.