Specifications
1-8
8753ES Option 011 Specifications and Characteristics
Instrument Specifications
Table 1-5 Receiver
8753ES Option 011 Receiver
Description Specification Supplemental Information
Internally Generated Harmonics (Option 002 Only)
2nd Harmonic 16 MHz to 3 GHz
at +0 dBm input level < −15 dBc
at −10 dBm input level < −30 dBc, typ.
at −30 dBm input level < −45 dBc, typ.
3rd Harmonic 16 MHz to 2 GHz
at +0 dBm input level < −30 dBc
at −10 dBm input level < −50 dBc, typ.
at −30 dBm input level < −50 dBc, typ.
Harmonic Measurement Accuracy
16 MHz to 3 GHz ±1.5 dB
3 GHz to 6 GHz ±3 dB
Harmonic Measurement Dynamic Range
Standard −40 dBc, typ.
output at −10 dBm and input
at < −15 dBm
Minimum R Level
300 kHz to 3 GHz −35 dBm
3 GHz to 6 GHz −30 dBm
Input Crosstalk
300 kHz to 1 GHz −100 dB
1 GHz to 3 GHz −90 dB
3 GHz to 4.5 GHz −82 dB
4.5 GHz to 6 GHz −75 dB