Specifications

1-11
8753ES Option 011 Specifications and Characteristics
Instrument Specifications
Table 1-9 Receiver
8753ES Option 011 Receiver (A and B Channel)
Dynamic Accuracy (Specification)
For inputs A and B, accuracy of the test port input power reading relative to the reference input
power level.
Inputs: test port A and B
For test port powers > 50 dBm and < 0 dBm, magnitude dynamic accuracy is
0.02 dB + 0.001 dB/dB from the reference power, phase dynamic accuracy is
0.132 deg + 0.0066 deg/dB from the reference power.