Specifications
4-15
Hardkey/Softkey Reference
Analyzer Functions
phase-locking the VCO.
switches the counter to count the A14 fractional-N VCO
frequency at the node shown on the overall block diagram.
switches the internal counter off and removes the counter
display from the LCD.
toggles the channel coupling of stimulus values. With
(the preset condition), both channels
have the same stimulus values (the inactive channel takes
on the stimulus values of the active channel).
is used to set the frequency for power sweep and CW time
sweep modes. If the instrument is not in either of these
two modes, it is automatically switched into CW time
mode.
turns on a sweep mode similar to an oscilloscope. The
analyzer is set to a single frequency, and the data is
displayed versus time. The frequency of the CW time
sweep is set with in the stimulus menu.
this math function ratios channels 1 and 2, and puts the
results in the channel 2 data array. Both channels must be
on and have the same number of points. Refer to the
"Making Measurements" chapter in the user’s guide for
information on how to use this function to make gain
compression measurements.
displays both the current data and memory traces.
specifies whether or not to store the error-corrected data
on disk with the instrument state.
divides the data by the memory, normalizing the data to
the memory, and displays the result. This is useful for
ratio comparison of two traces, for instance in
measurements of gain or attenuation.
subtracts the memory from the data. The vector
subtraction is performed on the complex data. This is
appropriate for storing a measured vector error, for
example directivity, and later subtracting it from the
device measurement.
stores the current active measurement data in the
memory of the active channel. It then becomes the
memory trace, for use in subsequent math manipulations
or display. If a parameter has just been changed and the *
status notation is displayed at the left of the display, the
data is not stored in memory until a clean sweep has been
executed. The gating and smoothing status of the trace are
COUNTER: FRAC N
COUNTER: OFF
COUPLED CH ON off
COUPLED CH ON
CW FREQ
CW TIME
CW FREQ
D2/D1 to D2 on OFF
DATA and MEMORY
DATA ARRAY on OFF
DATA/MEM
DATA - MEM
DATA
→
MEMORY