Specifications
4-18
Hardkey/Softkey Reference
Analyzer Functions
activates both forward and reverse measurements of
selected calibration standards.
activates both forward measurements (reflection and
transmission) of the thru standard from the selective
enhanced response calibration menus.
activates both reverse measurements of the thru standard
S22/S12 from the S11/S21 selective enhanced response
calibration menus
has two functions:
• It shows the current sequences in memory. To run a
sequence, press the softkey next to the desired
sequence title.
• When entered into a sequence, this command performs
a one-way jump to the sequence residing in the
specified sequence position (SEQUENCE 1 through 6).
jumps to a softkey position, not to a
specific sequence title. Whatever sequence is in the
selected softkey position will run when the
command is executed. This command
prompts the operator to select a destination sequence
position.
finishes one-port calibration (after all standards are
measured) and turns error correction on.
finishes two-port calibration (after all standards are
measured) and turns error correction on .
finishes the transmission portion of the enhanced response
calibration.
finishes all the load standards when the cal kit defines
more than one load standard.
finishes all the open standards when the cal kit defines
more than one open standard.
finishes all the short standards when the cal kit defines
more than one short standard.
finishes response and isolation calibration (after all
standards are measured) and turns error correction on.
finishes the transmission portion of the enhanced response
calibration.
terminates the sequencing edit mode.
finishes TRL/LRM two-port calibration (after all
standards are measured) and turns error correction on.
DO BOTH FWD + REV
DO BOTH FWD THRUS
DO BOTH REV THRUS
DO SEQUENCE
DO SEQUENCE
DO SEQUENCE
DONE 1-PORT CAL
DONE 2-PORT CAL
DONE FWD ENH RESP.
DONE LOADS
DONE OPENS
DONE SHORTS
DONE RESP ISOL'N CAL
DONE REV ENH RESP.
DONE SEQ MODIFY
DONE TRL/LRM